Carl Zeiss Axiolab A
3-16 B 40-015 e 06/99
3.3.1
Attachment of Pol Rotary Stage
Fig. 3-13 Pol rotary stage- setting of specimen mount and stop
The stage clips (3-13/3) are used to mount the specimen on the Pol rotary stage (3-13/8). They are
inserted into two of the three drilled holes (3-13/9) and, depending on the size of the specimen carrier or
the search area, used diagonally, opposite to each other or parallel.
The Pol object guide (3-13/11) (accessory 473325-0000-000) provides a more convenient possibility. It is
inserted in the drilled holes (3-13/7) and secured using the screw (3-13/10). For this purpose, use the
ball-headed screwdriver (3-13/12) supplied. The plastic rings (3-13/1) increasing the knob diameter (3-
13/4) for convenient operation from the side are particularly beneficial for screening.
To obtain a maximum screening area with the 24 × 48 specimen carrier size, mount the
specimen as shown in Fig. (3-13/2).