Zeiss N HBO 103 Specifiche Pagina 15

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 20
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 14
Microscope Illuminating System Start-Up
HBO 100 HBO 100 Carl Zeiss
B 46-0063 e 03/06 13
3.2 Start-Up of HBO 100
Microscope Illuminating
System
1 Connector for HBO 100 illuminator
2 Power connector
Fig. 5 Rear of transformer for HBO 100
1 LED LAMP
2 LED TEMP
3 LED SAFETY
4 LED +12V
5 Reset key
6 Display of operating hour meter
7 POWER switch with integrated power indicator LED
Fig. 6 Front of transformer for HBO 100
Do not look directly into the lamp or its
reflected light.
Procedure:
Install the HBO 100 microscope illuminator
on the microscope (see Operating Manual
of microscope).
Connect the microscope illuminator to the
transformer as described in the supplied
operating instructions:
Fig. 5
Connect the transformer to the power
outlet using the provided power cable.
Switch on the microscope illuminator using
the power switch at the front of the
transformer; the lamp will ignite
automatically:
Fig. 6
Function of the luminescence diodes:
"POWER" Lights when instrument is switched
on.
"LAMP" Lights when the lamp has been
ignited and is lighting.
"TEMP" Lights when the temperature
inside the transformer is within the
permissible range.
"SAFETY" Lights when the safety circuit of
the lamp housing is closed.
"+12V" Lights when the additional voltage
of the transformer is within the
permissible range.
Operating hour meter/"Reset" key:
Pressing the "Reset" key will reset the
operating hour meter to "0". This must be
done after every lamp change. To avoid
resetting the meter by mistake, the key can be
actuated only with an appropriate tool (e.g. a
small screwdriver).
Vedere la pagina 14