OPERATION
LSM 510 Quick start
B 40-051 e 07/98 4-9
4.3.3 Turning the Lasers On
(1) Click on the <Acquire> button from the LSM 510 tool bar.
(2) Click on the <Laser> button.
(3) You will see a Laser-Control menu with a list of available lasers. Using the mouse, click on the
laser(s) which has the appropriate wavelength to excite the dyes labeling your specimen.
(4) In the case of the argon and UV laser, click on the <Standby> button first.
Warming Up appears and, when the warming-up phase is finished, the Ready message. Then
click on the <On> button: the laser is switched on.
(5) Use Power[%] scrollbar to set the required laser power in %.
(6) In the case of the HeNe laser, click on the <On> button directly.
(7) Close menu.
Fig. 4-8
Fig. 4-9