OPERATION
Carl Zeiss Description of components and controls Axio Observer
68 B 46-0111 e 12/06
Manual iris stop slider for reflected light and
manual FL attenuator
The lever (4-8/4) on the slider is used to open
and close (lower position) the iris stop.
The two centering 3 mm Allen screws (4-8/2
and 3) can be used to center the stop in the
optical path.
The FL attenuator is used to attenuate the light
in the fluorescence path when using the
HBO 100 (slot A).
The FL attenuator has six labeled positions
which are selected by turning the control wheel.
Motorized iris stop slider for reflected light
and motorized FL attenuator
The motorized slider and motorized FL
attenuator are not compatible with the A1
stand.
The stop is opened or closed by pressing the
appropriate button on the slider.
The motorized (or manual) FL attenuator is used
to attenuate the light in the fluorescence path
when using the HBO 100 (slot A).
The FL attenuator has six positions which can be
selected in forward or reverse order by pressing
buttons (
4-9/1) and (4-9/2).
Drive knobs for controlling XY positioning of the mechanical stage (
4-1/13), (4-2/15), (4-3/14) or
object guide (if the specimen stage 250x230 with object guide is fitted)
Upper drive knob: movement in the Y-direction.
Lower drive knob: movement in the X-direction.
Fig. 4-8 Iris stop slider, manual
423646
Fig. 4-9 FL attenuator, motorized